Important Keys
7
8
9
0
Number
4
5
6
C
Keys
START
1
2
3
STOP
Alarm Light
Infusing Light
Setting the Pump
1. Press
(On/Off).
2. Press button on top of door to open.
3. Insert tubing into pump.
• Push open tube pincher A
• Place tubing in tube pincher A
• Insert tubing along channel B
• Push tubing into Air In Line Detector C
4. Close Door.
950853-Rev 06/02
5. Is this a new therapy?
Yes, No, OK, etc.
• Yes
• No
On/Off
(Power)
programmed infusion.
Alarm Silence
6. Prime set?
Start/Stop
• Yes
• No
Continuous Mode
1. Enter the rate using the number keys.
2. Press VTBD
3. Enter the bag volume using the number keys.
4. Press OK
A
B
5. Press START
C
: clears current infusion data.
: returns pump to last
: follow prompts to prime tubing.
: begin programming infusion.
.
.
.
Piggyback Mode
1. Press OPTS
until PIGGYBACK is displayed
2. Press ON
.
3. Press END
.
4. Press PIGY
.
5. Enter Rate/VTBD or VTBD/Time as in
Continuous Mode.
6. Press START
.
Changing the Rate
1. While infusing,
enter the new rate
using the number
keys.
2. Press OK
.
NOTE:
If the OK key is not pressed within 10 seconds
of the change, the infusion will continue at the
original rate.
Changing the VTBD
1. Stop infusion.
2. Press VTBD
.
3. Press
.
4. Enter the new
volume to be
delivered.
5. Press OK
.
6. Press
.