Page 1
(PN 217400). Ensure the frit is not damaged after pressing it into the vessel. NOTE CEM recommends replacing the polyethylene frit after ~25 peptides or 500 couplings. The frit should also be replaced in the event of a vessel clogging error or if the vessel fails a drain test.
Page 2
2. Select the Wash operation and change the volume to 10 mL and drain time to 10 seconds. NOTE If the reaction vessel fails to completely drain after 10 seconds, replace the frit. Replacement Parts Replacement parts can be ordered by visiting the Liberty Blue Replacement parts list on the CEM website at http://www.cem.com/liberty-blue-consumables.html. Part Part Number...