External Leak Test; Subjective Breathing Test - Aqua Lung GLACIA Service & Repair Manual

Second stage regulator
Table of Contents

Advertisement

16

EXTERNAL LEAK TEST

1.
After disconnecting the regulator from the flow bench, connect it to a scuba cylinder filled to approxi-
mately 3,000 psi. Open the cylinder valve to repressurize the regulator, and submerge the entire
system in a test tank of clean water.
2.
Observe any bubbles arising from the submerged regulator over a one minute period. The recom-
mended time is necessary due to slower bubble formation that occurs in smaller leaks. Bubbles indi-
cate a leak, which requires that the system must be disassembled at the source to check sealing sur-
faces, assembly sequence and component positioning in order to correct the problem(s).
NOTE: Extremely small leaks may be better detected by applying a soap solution or Snoop™ to the
leak area. Bubble streams will indicate the source of the leak. Before disassembling to correct any
leaks, rinse the entire regulator thoroughly with fresh water and blow out all residual moisture with
filtered, low-pressure (50 psi) air. Disassemble and remedy the problem, referring to " Table 1 -
Troubleshooting ."

SUBJECTIVE BREATHING TEST

1.
Depress the purge cover fully to ensure that an adequate volume of air needed to clear the second stage
flows through the mouthpiece. Then, inhale slowly but deeply from the mouthpiece. A properly ser-
viced and adjusted regulator should deliver air upon deep inhalation without excessive inhalation
effort, freeflow, or "fluttering" of the second-stage diaphragm. When exhaling, there should be no
fluttering or sticking of the exhalation valve. If any of these problems occur, refer to Table 1 -
Troubleshooting.
This concludes annual service procedures for the Glacia Second Stage Regulator.
© 2000 Aqua Lung America, Inc.
Glacia Second Stage Service & Repair Manual

Advertisement

Table of Contents
loading

Table of Contents