DISASSEMBLY PROCEDURES 3b. If the corrugated hoses are attached by clamping bands, use a small blade screwdriver and loosen the bands. NOTE: Before performing any disassembly, refer to the exploded parts drawing, which references all mandatory replacement parts. These parts should be replaced with new, and must not be reused under any circumstances –...
Page 5
Mistral Second Stage Technical Maintenance Manual 12. Place the handle of the seat extraction tool (p/n 109437) NOTE: Aqualung no longer requires directly over the poppet bearing (12) in the top center of the mandatory replacement of the exhaust valve body (11), and press downward until the poppet valves.
REASSEMBLY PROCEDURE 15. When the crown (16) has been unthreaded from the inlet fi tting (19), carefully insert the pin of the seat extraction tool into the opposite end of the inlet fi tting and through the NOTE: Before performing any reassembly, it center of the crown.
Page 7
Mistral Second Stage Technical Maintenance Manual 4. Using the crown adjusting tool (p/n 125727), engage the 8. Stand the poppet (14) on its head inside the top of the inlet slotted head of the crown (16) and drive it in clockwise as fi...
Page 8
CAUTION: Prior to adjusting and testing the Mistral second stage regulator, the acompanying fi rst stage must be correctly serviced and adjusted to a stable medium 12. Hold the valve body assembly (8-15) stationary and pressure of 135 ±5 psi and fully tested.
Page 9
Mistral Second Stage Technical Maintenance Manual Properly position the face of the adjusting tool (p/n 125727) on the 21. Fit 4 new locknuts (7) into the inhalation box, then tighten top of the sealing ring (4). Using a ¼" nut driver, tighten or loosen the screws (1) using a 3 mm allen key until snug.
FINAL TESTING PROCEDURES b. Align the mold lines on the corrugated hose and box. Install the “R” clamping band so that the lugs are in the same line as the fi rst stage body. Using a small NOTE: Before performing the following blade screwdriver, tighten all clamping bands.
Page 11
Mistral Second Stage Technical Maintenance Manual Table 1: Troubleshooting Guide SYMPTOM POSSIBLE CAUSE TREATMENT 1. Refer to First Stage Troubleshooting 1. MP is too high. Guide 2. LP seat (14) is worn or damaged. 2. Replace the seat 3. The crown (15) is not correctly adjusted.
Page 12
Table 2: List of Tools and Service Kits PART # DESCRIPTION APPLICATION MP Pressure Gauge 0-400 psi MP Pressure Gauge 0-400 psi 111610 Checking medium pressure Crown and Lever Adjusting Tool Crown and Lever Adjusting Tool 125727 Adjust crown (16) and lever (10) Seat Extraction Tool Seat Extraction Tool 109437...
Page 13
Mistral Second Stage Technical Maintenance Manual Table 3: Recommended Cleaners and Lubricants LUBRICANT/CLEANER APPLICATION SOURCE Aqua Lung, PN 820466, or Lubrication Technologies ® ® ® Christo-Lube MCG 111 MCG 111 All o-rings 310 Morton Street Jackson, OH 45640 (800) 477-8704...
When it comes to issues of nitrox safety and compatibility, the concerns lie primarily with the regulator’s fi rst stage as it is subjected to high inlet pressures. High inlet pressures lead to adiabatic compression or heating of the gas. The Aqua Lung or Apeks regulator product described in this manual, when properly cleaned and assembled, is authorized for use with enriched air nitrox (EAN) that does not exceed 40% (EAN 40).