Specifications - Sony UP-DR80MD Instructions For Use Manual

Digital
Hide thumbs Also See for UP-DR80MD:
Table of Contents

Advertisement

Specifications

Power Requirements
100 V to 240 V AC, 50/60 Hz
Input Current
3.4 A to 1.4 A
Operating Temperature
5 °C to 35 °C (41 °F to 95 °F)
Operating Humidity
20% to 80% (no condensation allowed)
Operating Pressure
700 hPa to 1,060 hPa
Storage and Transport Temperature
–20 °C to +60 °C (–4 °F to +140 °F)
Storage and Transport Humidity
20% to 80% (no condensation allowed)
Storage and Transport Pressure
700 hPa to 1,060 hPa
Dimensions
Approx. 317 × 425 × 207 mm (w × d × h)
Weight
Approx. 11.5 kg
Printing system Dye transfer sublimation thermal printing
Thermal head
11.83 dot/mm, 2,560 dots (301 dpi)
Gradations
8-bit (256-level) processing each for
yellow, magenta, and cyan
Maximum print size
A4 size: 202 × 287 mm
Letter size: 208 × 269 mm
Picture elements
A4 size: 2,392 dots × 3,400 dots
Letter size: 2,464 dots × 3,192 dots
Printing time
Approximately 76 seconds (A4 size)
Approximately 72 seconds (Letter size)
Interface
Hi-Speed USB (USB 2.0)
Accessories supplied
Paper holder (pink) (1)
Paper holder (blue) (1)
Cleaning ribbon (1)
USB cable 1-824-211-41 (SONY) (1)
Before Using this Printer (1)
Quick Guide (1)
CD-ROM (1)
Software License Agreement (1)
Service Contact List (1)
Information for Customers in Europe (1)
Optional accessories
Ink ribbon/paper roll
Self-Laminating Color Printing Pack
Self-Laminating Color Printing Pack
Wireless Print System
1)
As of July 2017, this system can be used in the EU and EFTA, the
United States, Canada, and Japan.
Design and specifications are subject to change without
notice.
26
Specifications
UPC-R80MD
UPC-R81MD
1)
UPA-WU10
Medical Specifications
Protection against electric shock:
Class I
Protection against harmful ingress of
water:
Ordinary
Degree of safety in the presence of a
flammable anesthetic mixture with air
or with oxygen or nitrous oxide:
Not suitable for use in the presence of
a flammable anesthetic mixture with
air or with oxygen or nitrous oxide
Mode of operation:
Continuous
Notes
• Always verify that the unit is operating properly before use.
SONY WILL NOT BE LIABLE FOR DAMAGES OF ANY
KIND INCLUDING, BUT NOT LIMITED TO,
COMPENSATION OR REIMBURSEMENT ON
ACCOUNT OF FAILURE TO PRINT CONTENT OF ANY
TYPE OR LOSS OF DATA DUE TO FAILURE OF THIS
UNIT OR ITS PRINTING MEDIA, ASSOCIATED
SOFTWARE, EXTERNAL STORAGE, OR OTHER
EXTERNAL DEVICE.
• SONY WILL NOT BE LIABLE FOR DAMAGES OF ANY
KIND INCLUDING, BUT NOT LIMITED TO,
COMPENSATION OR REIMBURSEMENT ON
ACCOUNT OF THE LOSS OF PRESENT OR
PROSPECTIVE PROFITS DUE TO FAILURE OF THIS
UNIT, EITHER DURING THE WARRANTY PERIOD OR
AFTER EXPIRATION OF THE WARRANTY, OR FOR
ANY OTHER REASON WHATSOEVER.