Guidance and manufacturer's declaration – electromagnetic immunity
The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of
the UP-DR80MD should assure that it is used in such an environment.
IEC 60601
Immunity test
Electrostatic
±8 kV contact
discharge (ESD)
IEC 61000-4-2
±15 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
Voltage dips, short
0% U
interruptions and
(100% dip in U
voltage variations
for 0.5/1 cycles
on power supply
input lines
40% U
(60% dip in U
IEC 61000-4-11
for 5 cycles
70% U
(30% dip in U
for 25/30 cycles
(for 0.5 sec)
0% U
(100% dip in U
for 250/300 cycles
(for 5 sec)
Power frequency
30 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
a For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.
Compliance level
test level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
0% U
T
T
)
(100% dip in U
T
a
for 0.5/1 cycles
40% U
T
T
)
(60% dip in U
T
for 5 cycles
70% U
T
T
)
(30% dip in U
T
a
for 25/30 cycles
(for 0.5 sec)
0% U
T
T
)
(100% dip in U
T
a
for 250/300 cycles
(for 5 sec)
30 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, a relative
humidity of at least 30% is recommended.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
)
T
UP-DR80MD requires continued operation during
a
power mains interruptions, it is recommended that the
UP-DR80MD be powered from an uninterruptible
power supply or a battery.
)
T
)
T
a
)
T
a
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
5