8.5
Autosampler
Connection of sampler changer TW alpha plus
8.5.1
The sample changer TW alpha plus is connected to the RS232-2 (RS2) of the piston burette with cable TZ 3087.
The settings of the RS232-2 interface must be changed to 4800, No, 7, 2 (Fig. 177).
Fig. 177
The settings of the RS232-1 (4800, No. 8, 1) remain unaffected.
Connection of sample changer TW 7400
8.5.2
The sample changer TW 7400 plus is connected to RS232-2 (RS2) of the piston burette by cable TZ 3987.
The settings of the RS232-2 interface do not have to be changed. They can remain at 4800, No. 8.1.
8.6
Using software TitriSoft
General
8.6.1
The piston burette is connected to the PC via the RS232 or USB-1-B interface. Cables TZ 3097 and TZ 3091 can
be used via RS232-1 for the connection.
TitriSoft 3.15 or higher
8.6.2
When using the new software 3.15 or higher TitriSoft, the factory settings of the RS232-1 can be maintained.