Xylem SI Analytics TITRONIC 500 Operating Manual page 36

Piston burette
Table of Contents

Advertisement

118
The following calculation formulae are available for manual titration:
Titration formula
(ml1-B)*T*M*F1/(W*F2)
(B-ml)*T*M*F1/(W*F2)
(B*F3–ml*F1)*T*M/(W*F2)
(W*F2)/(ml-B)*M*F1)
EP1
The abbreviations used here have the following meaning:
ml:
Titration solution consumption in ml
B:
Blank value in ml. Mostly determined by way of titration
T:
Titer of the titration solution (e.g. 0.09986)
M:
Mol; mol- or equivalence weight of the sample (e.g. NaCl 58.44)
F1
Factor 1, Conversion factor
F2
Factor 2, Conversion factor
"Weight", weighed-in quantity in g or volume in ml
W
After selecting a formula, please confirm your selection with <ENTER>/<OK>.
The values of the individual parameters of the selected calculation formula can now be input one by one (Fig. 66
and Fig. 67).
Fig. 66
Fig. 67
Additional information
Formula for calculating the concentration of a
sample taking into account a blank value in
terms of ml.
Formula for calculating the concentration of a
sample taking into account a blank value in
terms of ml. Reverse titration (examples. CSB,
saponification number)
Formula for calculating the concentration of a
sample taking into account a blank value,
including a multiplicative factor. Reverse
titration.
Formula for calculating a titer (T) of a titration
solution.
Used to calculate the consumption in ml

Advertisement

Table of Contents
loading

Table of Contents