Head Type
Standard, full size, Rectangular Heads
3" to flow Freefall Heads
5" to flow (and "CF") Freefall Heads
Standard Pipeline Heads
6000 Series Combo Heads
3000 Series Combo Heads
Table1
Computation of coil gap by head type and minimum aperture dimension
Step Three
The detector must now be set up for the desired user options. Until you are completely familiar with
the operation of the detector, it is recommended that all the options are left at their default values.
Press the [CANCEL] key once to move out of the Configuration menu, and then the [ ] key enough
times so that the display shows "USER OPTIONS". Press the [ENTER] key once to move into the
User Options menu, and then set up your options are desired:
TRACKER:
ISC:
REJECT TIME:
MAN SETUP:
AUX RELAY:
ACCESS FERR:
REJ OUT RUN:
CAL TRIGGER:
*PV TEST:
*FAULT PV:
Exit the User Options menu by pressing the [CANCEL] key once.
LOMA Superscan Micro ISC
COMMISSIONING GUIDE
This option will only appear if photo-eye operation has been selected in the
configuration menu. Turn this ON or OFF as required (see the A-Z section
for an explanation).
Turn this ON or OFF as required. This option is explained in the A-Z and
Principles of Operation sections.
Leave this at the default option of ON to enable access to the reject timing
information in the Product Setup menu. Only set this to OFF once the reject
time has been correctly set, and you with to deny line operators access to it.
Turn this ON or OFF as required. This option is explained in the A-Z
section.
Turn this to ERROR or RUN as required. Refer to User Options in the A-Z
section for more information.
This entry must be set to yes, in order that a ferrite tune can be performed -
see later.
Turn this ON or OFF as required. Refer to User Options in the A-Z section
for more information.
Turn this ON or OFF as required. Refer to User Options in the A-Z section
for more information.
Set to YES or NO as required. YES will enable the Performance Validation
system. Refer to the A-Z section and the Performance Validation system
section in Principles of operation for more information.
Set to YES or NO as required. YES will cause the fault relay to operate if a
Performance Validation system test is not actioned or completed in the time
allowed.
A GUIDE TO COMMISSIONING
Coil Gap
0.8 x Minimum Aperture Dimension
48mm
76mm
0.75 x Aperture Dimension
76mm
0.8 x Minimum Aperture Dimension
5