Page 2
1. General Information This digital multimeter is designed and manufactured in compliance with IEC-61010 safety requirements on electronic measuring instruments and hand-held digital multi-meters. It is compliant with IEC-61010 requirements pertaining to 600V CAT IV, 1000V CAT.Ⅲ and requirements on pollution degree 2.
Page 3
protective ring. When switching to another measuring range, be sure that test lead has already been taken off the measured circuit. For all DC functions, to prevent potential electric shock as a result of incorrect reading, please first use AC functions to check the absence of any AV voltage.
Page 4
Double insulation protection Fuse High voltage warning over-voltage protection CAT. III 1000 V over-voltage protection CAT. IV 600 V 1.1.4 Maintenance practices for safety The operator must first pull out the test lead when the meter's case is opened or the battery cover is dismantled. ...
Page 5
2. A Schematic Diagram for the Meter This meter is a hand-held digital multi-meter with the function of displaying True RMS. it is a large-screen LCD unit with backlight and illumination light functions so that the user can easily recognize reading.
Page 6
Table.1 Symbols Symbol Description Battery under voltage indicator/ Low Battery To avoid electric shock or personal injury as a result of incorrect reading, promptly replace the battery when the battery under voltage indicator appears. Auto power off indicator High voltage warning Negative input polarity indicator Input voltage AC Input voltage DC...
Page 7
Farad, the unit of capacitance. Millifarad, 1x10 or 0.001 farad. mF,F, nF F: Microfarad, 1x10 or 0.000001 farad. nF, 1x10 or 0.000000001 farad. 2.3 Description of functional keys Description of functions SEL keys, e.g. Temperature measurement position: ℃ mode or ℉ mode.
Page 8
2.5 Accessories 1.Operation Manual 2. Test lead X 1 pair 3. K-Type thermocouple 3. Operational guidelines 3.1 Normal operation 3.1.1 Hold mode In the hold mode, the reading can be maintained on the display unit. Changing the measurement function position or pressing the key. Hold again to exit the hold mode Hold mode: entry and exit 1.
Page 9
To avoid any electric shock and/or damage to the meter, do not attempt a voltage measurement if the voltage (valid value) is 1,000V for DC current or 750V for AC current. To avoid any electric shock and/or damage to the meter, don't attempt to impose between any public terminal and ground any voltage whose valid value is over 1,000V for DC current or 750V for AC current.
Page 10
measured and fully discharged all high-voltage capacitors. Ohm is the unit of electric resistance (). The measuring ranges of electric resistance of this meter are 600.0, 6.000k, 60.00k, 600.0k, 6.000Mand 60.00M, Measurement of electric resistance 1. Turn the rotary switch to the appropriate position. 2.
Page 11
the polarity of the test lead is reversed, the meter will display "OL". A normal diode still produces a forward voltage drop of 0.5V to 0.8V; the reverse bias voltage reading depend on the variation in electric resistance of other channels between two pen-shaped meters. 3.2.4 Beep continuity test To avoid the meter or the measured equipment from damage, do not attempt a beep continuity test unless the operator has...
Page 12
Note: The measurement of a large capacitance requires a given period of stabilization of reading. To avoid damage to the meter, the measurement of a capacitor with polarities requires much attention to its polarity. 3.2.6 hFE measurement To avoid any electric shock and/or damage to the meter, do not attempt a frequency measurement if the voltage is over 36V for DC current or AC current(valid value).
Page 13
function positions and measuring ranges. When a test lead is inserted into the current input socket, do not put the other end of the test lead in parallel connection with any circuit. The meter provides DC current measuring ranges as follows: 60A, 60.00mA, 600.0mA and 20.00A;...
Page 14
3.2.10 Live line test 1. Turn the rotary switch to the position Live. 2. Connect the test lead in red to Live input socket. 3. When a single test lead is inserted into jack L of power socket or approaches a live conductor, if the meter detects AC voltage, the meter's indicator for the intensity (high, medium and low) of induced voltage signal will be on and the beeper sounds alarms for different intensities.
Page 15
Input polarity indication: “” will automatically appear. Power: 4 x 1.5V AA battery Dimensions: 192 mm(L)×89mm(W)×55mm(H). Weight: approximately 330g (not included batteries or test leads) 4.2 Precision indicator Accuracy: ( reading + digit) The accuracy warranty will run for 1 year upon the ex-factory date. Reference conditions: ambient temperature is between 18℃...
Page 16
99.99KHz 0.01 KHz 999.9KHz 0.1 KHz 9.999MHz 0.001MHz Input voltage range: 200mV-10V ac valid value Overload protection: 600V DC/AC 4.2.4 Electric resistance Measuring range Resolution Accuracy 600 0.1 6k 1 ±(0.8% Reading + 3 digits) 60k 10 600k 100 6M 1k...
Page 17
measured resistance is no greater than 30, in approximately 40Ω-60Ω, The red indicator will be Overload protection: 600V DC/AC 4.2.7 Capacitor Measuring range Resolution Accuracy 0.001nF ± (4.0% Reading +30 digits) 60nF 0.01nF 600nF 0.1nF 6F ± (4.0% Reading +3 digits) 60F 10nF 600F...
Page 18
4.2.10 AC current Measuring range Resolution Accuracy 60mA 0.01mA ±(1% Reading + 3 digits) 600mA 0.1mA 10mA ±(1.5% Reading + 3 digits) Overload protection: protective tube for mA measuring range (FF630mA/250V) ; protective tube for 20A measuring range (10A/250V) . Maximal input current: mA position: 600mA DC/AC (valid value);...
Page 19
clean the meter's shell. Don't attempt the use of any abrading or chemical solvent. The dirty or damp input socket may affect reading. Steps for cleaning input sockets: Disenable the meter and pull all test leads out of the input socket. ...