Switching Amplifier; Analog Indicator With Transmitter; Commissioning; Adjusting The Alarm Signalling Unit - ABB FAM3200 Series Commissioning Instruction

Armored variable area purgemeter
Hide thumbs Also See for FAM3200 Series:
Table of Contents

Advertisement

6.5.2

Switching amplifier

Additional switching amplifiers are needed to operate the
alarm signaling units.
See chapter "Switching amplifier" on page 13 and the
"Ordering information" section of the data sheet for further
information.
Specifications
Power supply
230 V AC, +10 % / -15 %, 45 ... 60 Hz
115 V AC, +10 % / -15 %, 45 ... 60 Hz
24 V DC, +10 % / -15 %
Output
One or two switching relays with potential-free
changeover contacts
Switching capacity
Maximum 250 V, maximum 4 A, maximum
500 VA
Maximum permissible
Between switch amplifier and alarm signaling
cable length
unit: 300 m (984 ft)
Permissible ambient
-20 ... 60 °C (-4 ... 140 °F)
temperature range
Electrical connection
Screw terminals, maximum 2.5 mm
Type of assembly
35 mm top-hat rail in accordance with
EN 60715:2001
IP rating
IP 20 in accordance with EN 60529
Weight
Approx. 150 g (0.3 lb)
6.5.3

Analog indicator with transmitter

Specifications
Output signal
4 ... 20 mA, two-wire technology
Power supply
maximum 30 V DC
Power consumption
Maximum 30 mA
Connection type
Cable, 2-wire, brown (+) / blue (-), length
1.75 m (5.74 ft)
Ambient temperature
-20 ... 40 °C (-4 ... 104 °F)
range
Version
Angular position transmitter in two-wire
technology
1) Other cable lengths available on request.
2
(14 AWG)
1)
7

Commissioning

CAUTION
Risk of burns due to hot measuring media.
The device surface temperature may exceed 70 °C (158 °F),
depending on the measuring medium temperature!
Before starting work on the device, make sure that it has
cooled sufficiently.
During commissioning of the flowmeter, observe the following
points:
— The ambient and operating conditions (pressure,
temperature, power supply) must correspond to the
details on the name plate and the specifications.
— Open the shut-off valves slowly in order to prevent
pressure surges that can damage the flowmeter.
— For liquid measurement media, vent the pipeline if
necessary.
— For devices with alarm signalling units, set these to the
required switching points.
7.1

Adjusting the alarm signalling unit

The alarm signalling unit is designed as a proximity switch.
When inserted in the proximity switch, the metal tag initiates
the switching process. The switching point can be adjusted
with a screwdriver.
4
3
Fig. 6: Single alarm (example)
1 Metal tag 2 Alarm signalling unit (proximity switch)
3 Alarm adjustment 4 Indicator
1. Unscrew the housing cover.
2. Adjust the alarm setting to the required switching point
using a flat-bladed screwdriver.
3. Screw on housing cover.
7.2

Operating instructions

If there is a chance that safe operation is no longer possible,
take the device out of operation and secure it against
unintended startup.
1
2
G00745
FAM3200 | CI/FAM3200-EN Rev. B 13