Verification Tests
Functional Verification: Start Arm In Test
Description
The purpose of this test is to check the Start Arm In port. The "TRIG OUT" port of
the Command Module is used to send a Start Arm signal to the AFG.
Test Procedure
1 Reset the AFG:
2 Set up equipment as shown in Figure 2-2:
Figure 2-2 Start Arm In Test Setup
3 Send the following commands to the Command Module to output 0 V to the
4 Set up the AFG to output a 1 MHz sine wave, with an external Start Arm
24
*RST;*CLS
"Trig Out" port:
*RST
OUTP:EXT:STAT ON
OUTP:EXT:SOUR INT
OUTP:EXT:LEV 1
source:
FREQ 1E6;
:VOLT 4VPP
ARM:LAY2:SOUR EXT
INIT:IMM
Functional Verification: Start Arm In Test
Reset AFG and clear status registers
Set freq to 1 MHz
Set AFG amplitude
External Start Arm source
Initiate
Keysight E1445A Arbitrary Function Generator Service Guide