4.7.1
RS232 interface
The RS232 interface can be used to connect the CT Analyzer to a computer or to the optional CT SB2
switch box.
9-pole SUB-D connector, male
Figure shows outside view
onto the pins at the CT Analyzer!
6
nc
7
RTS (out)
8
CTS (in)
9
+ 5V
Do not connect!
For internal use only!
Housing: Shield (ground)
Figure 4-1: Pin assignment for RS232 remote control interface
9-pole (DB9) null modem or
crossover cable, 2 x female
Connections required:
1 2 3 4 5 6 7 8 shield
1 2 3 4 5 6 7 8 shield
Figure 4-2: Connection cable for RS232 remote control interface
4.7.2

USB interface

The USB interface can be used to connect the CT Analyzer to a computer. Communication via USB is
considerably faster than communication via RS232.
USB interface
(type B connector)
Figure 4-3: USB remote control interface (standard type B connector)
OMICRON
Technical data of the CT Analyzer
1
nc
2
RxD (data in / receive)
3
TxD (data out / transmit)
4
nc
5
GND (signal ground)