Temperature Calibration; Usb Power - Acoustic Research Ngara User Manual

Table of Contents

Advertisement

• Pressing SELECT increases the value of each selected digit by 1
• Pressing ENTER selects the next digit.
4. Once the expected level is entered, the user is given the chance to confirm or cancel calibration. "Y" confirms
entered value, "N" cancels calibration.
• Pressing SELECT toggles between "Y" and "N"
• Pressing ENTER accepts value.
This menu item is not accessible when in recording mode.

4.4.6 Temperature Calibration

Allows Ngara to calibrate its internal temperature sensor. The internal temperature sensor is used to monitor the
performance of the electronics within the weatherproof enclosure and it is not intended to replace weather station
data.
To calibrate, the following steps must be carried out -
2. Activate calibration mode by pressing ENTER
3. Ngara will wait until the expected level is entered using SELECT and ENTER
• Pressing SELECT increases the value of each selected digit by 1
• Pressing ENTER selects the next digit.
4. Once the expected level is entered, the user is given the chance to confirm or cancel calibration. "Y" confirms
entered value, "N" cancels calibration.
• Pressing SELECT toggles between "Y" and "N"
• Pressing ENTER accepts value.
This menu item is not accessible when in recording mode.

4.4.7 USB Power

The USB power menu item allows the toggling of the USB power setting to either Constant or Toggle.
To change the current timeout settings, the following steps must be carried out –
1. Activate menu item by pressing ENTER
2. Ngara will display current setting
3. Toggle setting using SELECT
Version — 25-7cd7566
Ngara
User Manual
Page — 28 of 80

Advertisement

Table of Contents
loading

Table of Contents