Installation; Operation; Configuration; Maintenance - Dräger VarioGard 2300 Instructions For Use Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 10

Installation

3
Installation
Details on the mechanical installation and the wiring diagram are
available in the VarioGard 23x0 installation instructions (order no.
90 33 469).
4

Operation

VarioGard 23x0 transmitter
For monitoring of the ambient
air, depending on the sensor
installed.
Three-coloured optical LED.
Contact points for operation
using a magnetic wand.
5

Configuration

Use the PC software VarioCal to prepare calibration and to adjust gas
settings.
12
6

Maintenance

Visual inspection
At least every 3 months to determine operational readiness.
Routine inspections
Dräger recommends having the system inspected by Dräger Service
at least once a year.
When necessary
Replace the sensors of the transmitters.
Calibration of transmitters
The calibration can be carried out with the magnetic wand (order no.
45 43 428) directly on the transmitter.
CAUTION
!
Always calibrate the zero-point before span.
Handling the magnetic wand:
Each transmitter has two contact
points labelled with
and
on the housing. Place the
magnetic wand on the contact
points for calibration.
Dräger VarioGard 2300 / 2320