Dräger VarioGard 2300 Instructions For Use Manual page 15

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 10
Status
Action
display
Carry out span adjustment:
Place the magnetic wand
Green flashes Maintenance
on the marking »
« for
2 seconds and then
remove it again.
Stop the calibration gas.
When the calibration has been completed, the
Remove the calibration
transmitter must be switched back to
adaptor VarioGard 23x0.
measurement mode using the magnetic wand.
Ensure that the
Place the magnetic wand on the marking »
transmitter is no longer
for more than 5 seconds and then remove it again.
being purged with
The calibrations that were successfully carried out
calibration gas.
will be accepted.
Dräger VarioGard 2300 / 2320
Output
Meaning
signal
Being adjusted.
mode
«
Maintenance
15