Types Of Oil; Technical Data; General Technical Data; Operating Conditions - Heraeus Combilabor CL-P 7 Operating Instructions Manual

Vacuum pump
Table of Contents

Advertisement

Available languages

Available languages

Operating instructions vacuum pump – CL-P Type 7
9

Types of oil

Use exclusively unalloyed compressor oil according to DIN 51506, lubricant class VC, ISO VG 32 (SAE 10) for normal surroun-
ding temperatures. For surrounding temperatures above 30°C (86°F) or operation at 60 hz oil according to ISO VG 68 (SAE 20)
must be used.
Original pump oil for oil change:
100 ml
Order No.: 645 02 694
250 ml
Order No.: 645 02 692
CAUTION!
Note following:
If oil change intervals are not adhered to, there is a risk of corrosion and damage to the pump housing!
Used oil must not reach the sewage system!
Observe the legal provisions on the handling of used oil!
10

Technical data

10.1 General technical data

Designation
Mains voltage:
Mains frequency:
Nominal suction power:
Ultimate pressure:
Speed:
Motor power:
Starting capacitor:
Sound pressure level:
Oil capacity:
Weight:
10.2 Storage and transport conditions
Temperature:
Relative air humidity:

10.3 Operating conditions

Temperature:
Air humidity:

10.4 Type plate

Value
100 – 110 at 50 Hz
100 – 120 at 60 Hz
230
50 / 60
4.0 / 4.8
< 10
Betriebsanleitung Vakuumpumpe - CL-P Typ 7
3000
3600
10.1 Technische Daten
0.1 / 0.12
6
Benennung
< 70
Netzspannung:
60
Netzfrequenz:
Nennsaugvermögen:
5.75
Enddruck:
Drehzahl:
Motorleistung:
Anlaufkondensator
Schalldruckpegel
Ölfüllung
–15°C to + 40°C (5°F to 104°F)
Gewicht
10.2 Transport-, Lagerbedingungen
less than 100% rel. moisture
Temperatur
relative Luftfeuchte
10.3 Betriebsbedingungen
5°C to 40°C (41°F to 104°F)
Temperatur
Luftfeuchte
80% rel. humidity up to 31°C (88°F)
50% rel. humidity at 40°C (104°F)
10.4 Typenschild
Unit
V
Hz
m
3
/h
mbar
r/min. at 50Hz
r/min. at 60Hz
KW
μF
Wert
100 - 110 bei 50 Hz
db(A)
100 - 120 bei 60Hz
230
ml
50/60
4,0 / 4,8
kg
< 10
3000
3600
0,1 / 0,12
6
< 70
60
5,75
-15°C bis + 40 °C
kleiner 100% rel. Feuchte
+5°C bis + 40 °C
80% rel. Feuchte bis 31 °C
50% rel. Feuchte bei 40 °C
Heraeus Kulzer GmbH
COMBILABOR Vakuumpumpe / Vacuum pump
Bauart:
Fabrik Nr. / ID. No.:
Netzspannung, power supply:
Netzfrequenz, frequency:
Leistung:
Öl / Oil SAE 32
CE:
MADE IN GERMANY
Bild 4 Typenschild CL-P Typ 7
GB
- 9 -
Einheit
V
Hz
m3/h
mbar
U/min. bei 50Hz
U/min. bei 60Hz
KW
µF
db(A)
ml
kg
Grüner Weg 11
D-63450 Hanau
Germany
CL P Typ 7
82 1 33113
230 V ; AC; 1/N/PE
50/60 Hz
0,1 KW
0,06 ltr.

Advertisement

Chapters

Table of Contents
loading

Related Products for Heraeus Combilabor CL-P 7

This manual is also suitable for:

660 02 450660 04 332

Table of Contents