Errors and causes

The following possible causes should help for trouble-shooting. If an error can not be eliminated,
the responsible service agent (see paragraph 11 Service) must be informed.
Motor does not start
Check socket and fuse at the casting machine / ceramic furnace.
Check mains connection. (cable and plug).
Check starting capacitor (see paragraph 5.1 Elements of device, pos. 4). (service)
GB
Check oil level and change oil if required.
- 8 -
Jamming of pump / motor does not pull through
Pump has run without oil –> inform service (see paragraph 11 Service).
Pump starts with difficulties / runs too slowly
contaminated oil –> change oil.
Too much oil filled in –> drain excess oil.
Incorrect type of oil –> change oil.
Pump has not been operated for a longer time (weeks or months) –> pump with closed intake.
Let pump run for approx. 5 minutes, change oil if required, look for corrosion. (corrosion particles in the oil?)
Starting capacitor (see paragraph 5.1 Elements of device, pos. 4) defective. (service)
Insufficient vacuum
Intake (suction) line leaky –> check clamps and tube.
Check oil level and fill up if required.
Filter element (see paragraph 5.1 Elements of device, pos. 2) clogged –> exchange filter element.
Pump gets unusually hot
normal working temperature approx. 80°C to 90°C (176°F to 194°F).
Blocked cooling air supply –> do not block fan openings.
Too much oil filled in –>drain excess oil.
Filter element (see paragraph 5.1 Elements of device, pos. 2) clogged –> exchange filter element.
Oil mist resp. loss of oil
Too much oil filled in –> drain excess oil, exchange filter element (see paragraph 5.1 Elements of device, pos. 2).
Pump was tilted with oil filled in –> exchange filter (see paragraph 5.1 Elements of device, pos. 2).
Incorrect type of oil –> change oil, exchange filter (see paragraph 5.1 Elements of device, pos. 2).
Filter element (see paragraph 5.1 Elements of device, pos. 2) used –> exchange filter.
If errors can not be eliminated: inform service, addresses can be found in paragraph 11 Service.