ARS CALIBRATION PROCESS
1.
Press the power button to turn on the ALERT J4X breath alcohol tester.
2.
CAL is displayed when the tester is overdue for a calibration. Press the power
button quickly in this sequence:
(CLICK) – (CLICK) – (PRESS AND HOLD UNTIL IT FLASHES)
3.
The backlight flashes blue; count the flashes for the next step.
4.
Press the power button as many times as the backlight flashed.
UEt is displayed.
5.
Press and hold the power button until 001 is displayed.
6.
Insert the tip of the square mouthpiece from the simulator into the sensor inlet of
the tester.
7.
Blow into the round mouthpiece of the simulator. Keep blowing until the Wait icon
appears.
4 samples are required. 001, 002, 003 and 004 indicate the current sample
number.
When the test is accepted and the calibration is complete, the screen will blink and
return to a Ready condition.
NOTE: If the calibration fails on numerous attempts, the sensor may need to be
replaced. Contact an authorized service provider for return instructions.