General Specifications; General Characteristics - Keysight Technologies N5261A User's And Service Manual

Millimeter head controller
Hide thumbs Also See for N5261A:
Table of Contents

Advertisement

Installation & System Operation

General Specifications

Specifications for the test set are characteristic of the test set. A
functional certificate is only offered for the test set. A periodic
calibration of the test set is not required.

General Characteristics

Environmental:
Temperature
Altitude
Power Requirements:
Nominal Frequency Range
Nominal Voltage Range
N5261A or N5262A Power
Weight and Dimensions:
Net Weight
Dimensions
This product is designed for use in Installation Category II and Pollution
Degree 2.
Required Conditions for Accuracy Enhanced Measurement
Accuracy-enhanced (error-corrected) measurements require the ambient temperature of the
N5261/62A to be maintained within ± 1 °C of the ambient temperature at calibration.
The instrument can safely operate in a relative humidity of 80% for temperatures to 31 degrees
C, decreasing linearly to 50% relative humidity at 40 degrees C.
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
Keysight N5261/62A User's & Service Guide
20 to 30 °C
3,000 meters (9,842 ft)
50/60 Hz
100/120/220/240 Vac
350 Watts (Max.)
N5261A 10 kg (22 lb)
N5262A 11 kg (24.2 lb)
Height: 18 cm (7.1 in)
Width: 42.5 cm (16.75 in)
Depth: 42.5 cm (16.75 in)