Guidance and Manufacturer's Declaration – Electromagnetic
Emissions
The G210 is intended for use in the electromagnetic environment specified below. The customer or the user of
the G210 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic
Immunity
The G210 is intended for use in the electromagnetic environment specified below. The customer or the user of
the G210 should assure that it is used in such an environment.
IMMUNITY Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Compliance
The G210 uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The G210 is suitable for use in all establishments other than
Class A – Complies
domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used
Class A - Complies
for domestic purposes.
Class A - Complies
IEC 61326-1
Test level
+ 4 kV contact
+ 2, 4, 8 kV air
+/- 1.0kV Burst Potential -
Mains
+/- 0.5kV Burst Potential -
Signals/Control
Line-Line: 0.5kV
Combination Wave (1.2μS
x 50μS Voltage, 8μS x 20μS
Current)
Line-Earth: 0.5kV & 1.0 kV
Combination Wave (1.2μS
x 50μS Voltage, 8μS x 20μS
Current)
100% Reduction; 0.5-1
50Hz Period Cycles; 0° &
180° Sync Angle
30% Reduction; 0.5s
Duration; 25 50Hz Period
Cycles; 0° & 180° Sync
Angle
100% Reduction; 5s
Duration; 250 50Hz Period
Cycles; 0° Sync Angle
Electromagnetic environment - guidance
Compliance level
environment - guidance
Floors should be wood,
concrete or ceramic tile.
+ 2, 4 kV contact
If floors are covered with
+ 2, 4, 8 kV air
synthetic material, the relative
humidity should be at least
30 %.
+/- 1.0kV Burst
Potential - Mains
+/- 0.5kV Burst
Potential - Signals/
Control
Line-Line: 0.5kV
Combination Wave
(1.2μS x 50μS Voltage,
8μS x 20μS Current)
Line-Earth: 0.5kV & 1.0
kV Combination Wave
(1.2μS x 50μS Voltage,
8μS x 20μS Current)
100% Reduction; 0.5-1
50Hz Period Cycles; 0°
& 180° Sync Angle
30% Reduction; 0.5s
Duration; 25 50Hz
Period Cycles; 0° &
180° Sync Angle
100% Reduction; 5s
Duration; 250 50Hz
Period Cycles; 0° Sync
Angle
Safety
Electromagnetic
2
9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the G210 InviCell and is the answer not in the manual?

Questions and answers

Related Products for K-Systems G210 InviCell

This manual is also suitable for:

Invicell standard signiphyInvicell plus signiphy

Table of Contents

Save PDF