Replacing the Origio Gas line
Filter
Remove all samples from the chambers before
replacing the filters.
1.
To access to the filter compartment, push
the safety lock lever (1) and lift the upper
part of the incubator containing the
chambers.
2.
Fasten the connectors to the inlet and
outlet of the Origio Gas Line Filter (2).
3.
Ensure the direction of flow shown on the
Origio Gas Line Filter label (3) is consistent
with the flow direction (4) shown inside the
filter compartment.
4.
Place the Origio Gas Line Filter in the filter
holder with the label facing up (5).
5.
Lower the upper part of the incubator.
CAUTION: Do not use the unit without a genuine Origio Gas Line Filter.
CAUTION: Insert filter with the filter label facing up. If the filter is not placed correctly, it
can cause excessive gas consumption.

Disposal of the Origio Gas line Filter

CAUTION: Contamination Hazard
As the filter may have been used for processing and treating infectious substances, it
might be contaminated. The used Gas Line Filter should be placed in a sealed plastic bag
and labeled as biohazard material, then disposed of according to local requirements.
1
2
4
5
3
Service
13
49

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the G210 InviCell and is the answer not in the manual?

Questions and answers

Related Products for K-Systems G210 InviCell

This manual is also suitable for:

Invicell standard signiphyInvicell plus signiphy

Table of Contents

Save PDF