Indication
Error description
E04: Temperature Error Unknown Barcode
The environmental temperature is above or
below the operating range of the analyzer.
Solution
Move to an area in the acceptable temperature
range for the test device, and perform the test.
Do not heat or cool the analyzer artificially.
E05: Communication Error
The communication between analyzer and
barcode or printer is failed.
Solution
Connect again between analyzer and external
device. If the error continues after turning
ON/OFF the analyzer, please contact SD
BIOSENSOR, Inc.
E06: Out of total hemoglobin
The measured total hemoglobin is out of the
range of 7 to 23g/dL.
Solution
This error occurs with a specimen known to
have total hemoglobin in the abnormal range.
If the error continues after turning ON/OFF the
analyzer, please contact SD BIOSENSOR, Inc.
Result: Extremely Low C line
The test is invalid.
Solution
Re-test with the new test device and the new
patient's specimen. If the error continues after
turning ON/OFF the analyzer, please contact
SD BIOSENSOR, Inc.