How To Run The Calibration Set Test; Chapter 06. Calibration Set - SD Biosensor F200 User Manual

Standard
Table of Contents

Advertisement

CHAPTER
06. Calibration Set
Calibration Set Test
Calibration set test is the required function that ensures optimal performance by
checking the designated method.
When to use calibration set
· Whenever the analyzer is powered on.
· When you drop the analyzer.
· Whenever you do not agree with your result.
· When you want to check the performance of an analyzer and test device.
Precautions and warnings
· The calibration set test cannot be replaced with the control test.

How to Run the Calibration Set Test

1. To run the calibration set test, select the 'Calibration' menu. Enter operator ID
before the staring calibration set test.
The operator ID can be entered into the analyzer with touch panel
or USB keyboard. And they can be entered with SDB barcode
reader as well which is purchased separately.
Note
51

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the F200 and is the answer not in the manual?

Table of Contents

Save PDF