CHAPTER
06. Calibration Set
Calibration Set Test
Calibration set test is the required function that ensures optimal performance by
checking the designated method.
When to use calibration set
· Whenever the analyzer is powered on.
· When you drop the analyzer.
· Whenever you do not agree with your result.
· When you want to check the performance of an analyzer and test device.
Precautions and warnings
· The calibration set test cannot be replaced with the control test.
How to Run the Calibration Set Test
1. To run the calibration set test, select the 'Calibration' menu. Enter operator ID
before the staring calibration set test.
The operator ID can be entered into the analyzer with touch panel
or USB keyboard. And they can be entered with SDB barcode
reader as well which is purchased separately.