Table 2: 2230-30-1 And 2230J-30-1 Specifications For Channels 1 And 2 - Keithley 2200 Series Specification And Performance Verification Technical Reference

Multichannel programmable dc power supplies
Hide thumbs Also See for 2200 Series:
Table of Contents

Advertisement

Specifications
Table 2: 2230-30-1 and 2230J-30-1 specifications for Channels 1 and 2 (cont.)
Parameter
Remote Sense Voltage Range
Voltage Resolution, nominal
√ Voltage Setting Accuracy with
Remote Sense
√ Voltage Setting Accuracy without
Remote Sense
Current Limit, nominal
Voltage Temperature Coefficient,
typical
Constant Current Operation,
nominal
Current Range
Current Resolution, nominal
√ Current Setting Accuracy
Voltage Limit, nominal
Current Temperature Coefficient,
typical
Meter, nominal
Current Readback Resolution,
nominal
Current Readback Accuracy
Current Readback Temperature
Coefficient, typical
Voltage Readback Accuracy
Voltage Readback Resolution,
nominal
Voltage Readback Temperature
Coefficient, typical
Line regulation – Voltage
Line regulation – Current
Load regulation – Voltage
4
Description
The total voltage from - terminal to + terminal may not exceed the rated full scale voltage +0.2V,
allowing a minimum of 0.1V/line for remote sense.
The Maximum voltage difference between a terminal and its associated sense line may not
exceed 1V.
The available overhead above full scale voltage to support remote sense is limited to 0.1
V/line. However, this widens as the output voltage decreases to the ultimate limit of 1 V/line.
Thus, the maximum difference specification controls operation except for the 2 V of output
below full scale output.
1 mV
+/-(0.03% of setting + 10 mV) at 25 °C ±5 °C with Remote Sense
+/-(0.05% of setting + 10 mV) at 25 °C ±5 °C without Remote Sense
Defined by Constant Current setting. Changeover is automatic.
+/-(0.03% of setting + 10 mV) per °C outside of the 25 °C ±5 °C standard range
The unit may be set to be a constant current over a range of voltages.
The output voltage may be adjusted from 0 A to 1.5 A.
1 mA
+/-(0.1% + 5 mA) at 25 °C ±5 °C.
0 to 30 V.
+/-(0.1% + 5 mA) per °C outside of the 25 °C ±5 °C standard range
Internal readback capability. Voltage and current are measured internally.
1 mA
±(0.01% of reading + 5 mA) at 25 °C ±5 °C
±(0.01% + 5 mA) per °C outside the 25 °C ±5 °C standard range
±(0.03% of reading + 10 mV) at 25 °C ±5 °C
1 mV
±(0.03% of reading + 10 mV) per °C outside the 25 °C ±5 °C standard range
Change over the full rated line voltage range: ±(0.01% + 3 mV)
Change over full rated line voltage range: ±(0.1% + 3 mA)
Change from 0 to 100% of full scale. Voltage: ±(0.01% + 3 mV)