Section 4. Parameter Setup

Entering and Exiting Setup Mode:

Quick Setup:
Quick Setup is recommended for customer use as this setup mode allows the non-
meteorological parameters to be modified and will not affect scale calibration.
Parameters with the
To enter the front-panel setup mode, perform the following steps:
1. Turn off the scale.
2. Press and hold the UNITS button.
3. Press and hold the ON/OFF button for approximately one second.
4. Release both buttons.
The scale will now be in the Quick Setup mode as indicated by "Ao XX".
Full Setup:
The Full Setup mode is recommended for authorized technician use only. To
enter the complete setup menu, perform the following steps:
1. Turn on the PC-500.
2. Turn the scale over and remove the battery compartment cover.
3. Press the calibration switch.
The scale should enter the Setup/Calibration mode as indicated by "div 2.5",
"div 5.0", or "div 10", (depending on the scale's resolution.)

Scrolling Through Parameters:

Pressing UNITS button will cause the scale to step to the next parameter on the list.

Changing a Parameter:

Once the desired parameter is displayed, it may adjusted. Press ZERO to step
through the parameter selections.

Saving Parameter Changes:

To exit the setup mode, press and release the UNITS button until the scale returns to
the weighing mode. At this time the parameter changes will be saved. Setup
information is not saved until the PC-500 exits the setup mode. A power failure while
in the Setup Mode will cause changes to be lost.

Changing Start up Units:

The scale will "power-up" with the same units that it was using when it was last turned
off.

Legal for Trade Restrictions:

The Legal for Trade mode disables the lb:oz mode and forces several parameters to
specific settings. These restricted parameters are indicated by an asterisk. Changes
to these parameters will increment the calibration audit counter.
symbol in section 5 are available in this menu.