7.1 Maintenance Schedule

The chair requires regular maintenance. Set up and follow
a maintenance schedule. The table at right represents
minimum intervals for maintenance.
Keep maintenance records. A sample maintenance record
sheet is provided on page 35.
When using maintenance products, follow the manufacturers'
directions and read the manufacturers' material safety data
sheets. You can purchase a recommended disinfectant from
your Ferno distributor or Ferno Customer Relations (page 2).

7.2 Disinfecting/Cleaning Restraints

Ankle Restraint: Place buckled restraint in a mesh bag
and launder in a washing machine using warm water, a
disinfectant soap, and gentle agitation. Hang the restraint to
dry. Do not put it in a dryer.
Chest and Lap Restraints: Disinfect and clean restraints as
directed in the restraint users' manual provided. Additional,
free users' manuals can be obtained from Ferno Customer
Relations (page 2).

7.3 Disinfecting/Cleaning Chair

To disinfect: Remove restraints and seat and backrest
panels or soft vinyl cover. Wipe all surfaces of the chair and
panels or cover with disinfectant. Follow the disinfectant
manufacturer's directions for application method and
contact time.
To clean: Wash panels or cover, and all parts of the chair,
with warm water and a mild detergent. Rinse with clean
water. Dry with a towel or allow to air-dry.

7.4 Cleaning Track and Belts

To ensure predictable belt performance, it is important to
keep the track and belts free of oil, grease, dirt, and debris.
1.
Hold the belts away from the track frame and remove
any debris.
2.
Clean both sides of the track frame, and both belts,
with SD-20 All-Purpose Cleaner and a clean cloth.
Follow instructions on the container.
3.
Dry track and belts thoroughly.
Note:

To order SD-20 Cleaner, call Spartan Chemical
Company at 1.800.537.8990.
© Ferno-Washington, Inc 234-3298-02 May 2009

7 - MAINTENANCE

Disinfecting Chair (this page)
Cleaning Chair (this page)
Cleaning Track System (this page)
Inspecting Chair (page 24)
Inspecting Track System (page 24)
Lubricating (page 25)
Improper maintenance can cause injury. Maintain
the chair only as described in this manual.
Disinfectants
phenolics, or iodines can cause damage. Use caution
when disinfecting and cleaning with products
containing these chemicals.
Water under high pressure, or steam, can penetrate
joints, flush away lubricant, and cause corrosion. Use
caution when cleaning moving parts such as joints and
hinges.
Using abrasive cleaning compounds or applicators
on the chair can cause damage. Do not use abrasive
materials to clean the chair.
Minimum
Maintenance Intervals
WARNING
Important
and
cleaners
Important
Important
Maintenance
containing
bleach,
23