Performance characteristics of metal process isolating
diaphragm
• As per IEC 60770
• Ambient temperature T
A
• Humidity φ = constant, in the range of 5 to 80 % rH
• Ambient pressure p
= constant, in the range of:860 to 1 060 mbar (12.47 to 15.37 psi)
A
• Position of measuring cell = constant, in range: horizontal ±1° (see also "Influence of the
installation position" section →  23)
• Zero based span
• Process isolating diaphragm material: AISI 316L (1.4435)
• Filling oil: synthetic oil polyalphaolefin FDA 21 CFR 178.3620, NSF H1
• Supply voltage: 24 V DC ±3 V DC
• Load: 320 Ω (at 4 to 20 mA output)
The smallest extended uncertainty of measurement that can delivered by our standards is:
• in the range 1 to 30 mbar (0.0145 to 0.435 psi): 0.4 % of reading
• in the range < 1 mbar (0.0145 psi): 1 % of reading.
→  23
Current output: min. 1.6 μA
The reference accuracy contains the non-linearity [DIN EN 61298-2 3.11] including the pressure
hysteresis [DIN EN 61298-23.13] and non-repeatability [DIN EN 61298-2 3.11] in accordance with
the limit point method as per [DIN EN 60770].
Device
% of the calibrated span to the maximum turn down
Reference accuracy
1)
PMP11
±0.5
PMP21
±0.3
1)
For devices with 0 to 10 V output, a non-linearity of up to max. 0.3 V can occur for signal values below
0.015 V.
Overview of the turn down ranges →  15
Measuring cell
–20 to +85 °C (–4 to +185 °F)
% of the calibrated span for TD 1:1
<1 bar (15 psi)
<1
≥ 1 bar (15 psi)
<0.8
Device
1 year
PMP11
±0.2
PMP21
≤2 s
Cerabar PMC11, PMC21, PMP11, PMP21
= constant, in the range of:+21 to +33 °C (+70 to +91 °F)
Non-linearity
±0.1
±0.1
5 years
±0.4
Non-repeatability
±0.1
±0.1
–20 to –40 °C (–4 to –40 °F)
+85 to +100 °C (+185 to +212 °F)
<1.2
<1
8 years
% of URL
±0.45