Operability; Plug-On Display Phx20 (Optional) - Endress+Hauser Cerabar PMC11 Technical Information

Pressure transducer with ceramic and metal sensors
Hide thumbs Also See for Cerabar PMC11:
Table of Contents

Advertisement

Plug-on display PHX20
(optional)
42

Operability

Devices with valve plug can be fitted with the optional local display PHX20.
Designation
Plug-on display PHX20, IP65
1)
Product Configurator, order code for "Accessories"
A 1-line liquid crystal display (LCD) is used. The local display shows measured values, fault messages
and information messages. The device display can be turned in 90° steps. Depending on the
orientation of the device, it is therefore easy to read the measured values.
Technical data
Display:
Digit height:
Display range:
Accuracy:
Electrical connection:
Power supply for display:
Voltage drop:
Conversion rate:
Damping:
Data backup:
Error message:
Programming:
Degree of protection:
Effect of temperature on display:
Electromagnetic compatibility
(EMC):
Permitted current load:
Ambient temperature:
Housing material:
Order number:
Cerabar PMC11, PMC21, PMP11, PMP21
4-digit, red LED display
7.62 mm; programmable decimal point setting
-1999...9999
0.2% of span ±1 digit
to transmitter with 4 to 20 mA output and elbow plug DIN 43 650, with
reverse polarity protection
not required, self-powered by the current loop
≤ 5 V (corresponds to load: max. 250 Ω)
3 measurements per second
0.3 to 20 s (configurable)
non-volatile EEPROM
• HI: Overranging
• LO: Underranging
via 2 buttons, menu-guided, scaling of display range, decimal point,
damping, error message
IP 65
0.1% / 10 K
Interference emission as per EN 50081, interference immunity as per EN
50082
max. 60 mA
0 to +60 °C (+32 to +140 °F)
Plastic Pa6 GF30, blue
Front screen made of PMMA, red
52022914
1)
Option
RU
Endress+Hauser