Charging of Battery
3�4�2
Charging with the DC Power
Cable
IMPORTANT
• If the power is supplied through the DC Power Cable with-
out inserting the Battery, the surface temperature of the DR
Detector may become high.
1
Make sure that the Battery is inserted to
the DR Detector.
2
Connect the wired connection connector
of the DC Power Cable to the wired con-
nection connector of the DR Detector.
• Charging starts.
Wired connection connector
DR Detector
Wired connection connector
• When charging starts, the Battery LED (green/
orange) on the DR Detector blinks.
• When charged more than 11 %, the Battery
LED (green) blinks.
Battery LED
(green/orange)
HINT
• The DR Detector will be charged when it is in the sleep mode
or turned off.
• The wired connection connector of the DC Power Cable
does not have an insertion direction.
• When inserting the DR Detector to the wall stand or table
for exposure, confirm the level of battery power with the im-
age processing controller.
• When the DR Detector is on, the charging time will be
slightly longer as it depends on the operation status.
• Connecting the DC Power Cable when the DR Detector is
off will turn the DR Detector on.
DC Power Cable