Konica Minolta AeroDR SKR 4000 Operation Manual page 17

Direct digitizer
Table of Contents

Advertisement

Table 1 Emissions
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
Class B
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
Complies
IEC 61000-3-3
Table 2 Immunity
Phenomenon
Electrostatic discharge
IEC 61000-4-2
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
wireless communications
IEC 61000-4-3
equipment
Electrical fast transients /
IEC 61000-4-4
bursts
Surges
IEC 61000-4-5
Conducted disturbances
IEC 61000-4-6
induced by RF fields
Rated power frequency
IEC 61000-4-8
magnetic fields
Voltage dips
IEC 61000-4-11
Voltage interruptions
IEC 61000-4-11
[NOTE]
U
is the a.c. mains voltage prior to application of the test level.
T
a The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz
to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
b E.g. 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz.
Dedicated medical facility environment
Dedicated medical facility environment
Dedicated medical facility environment
Basic EMC standard or test
method
Electromagnetic environment - guidelines
Immunity test levels
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
3 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
See the following table "Test specifications for Enclosure port
immunity to RF wireless communications equipment"
±2 kV input d.c. power port
±2 kV input d.c. power port
±1 kV signal input/output parts port
100 kHz repetition frequency
±0.5 kV, ±1 kV line-to-line
±0.5 kV, ±1 kV, ±2 kV line-to-ground
3 V
0.15 MHz - 80 MHz
6 V in ISM bands between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
30 A/m
0 % U
; 0.5 cycle
T
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % U
; 1 cycle
T
and
b
70 % U
; 25/30 cycle
T
Single phase: at 0°
b
0 % U
; 250/300 cycle
T
17