Welch Allyn ELI 280 Addendum User Manual page 14

Exercise stress option
Hide thumbs Also See for ELI 280:
Table of Contents

Advertisement

NIBP:
Allows entry of Systolic and Diastolic numeric values; displayed as mmHg in the display top
border .
NOTE: ECG waveform is not displayed during keyboard SBP and DBP value entry.
NOTE: The NIBP value is displayed and printed on ECG for 60-seconds after entered.
Abort:
Allows pre-exercise cancellation and a return to the Patient Information display.
ECG Display Settings
Touching the displayed ECG waveforms will open a menu allowing configuration of the lead layout, gain,
speed and filter settings. Select OK to save changes and return to ECG display. Select Cancel to return to the
ECG display without change.
ECG Printout
ECG printouts will use the ELI 280 ECG configuration and lead format from the resting ECG settings.
Cardiac stress exam ECG printouts include:
Patient demographics
Protocol Name (for treadmill tests)
Date/Time at start of exam
Total phase time (PRE, EXE, or REC)
Stage and stage time
Workload values
Heart rate
Target heart rate
Non-invasive blood pressure if less than 1-minute old
Auto/Manual/Peak indication
The ECG footer includes: serial number, site name, site number, cart number, software version, ECG speed,
gain, and filter settings.
USING THE ELI 280 EXERCISE STRESS OPTION
Page | 12

Advertisement

Table of Contents
loading

Table of Contents