4. Select the button for the type of exercise equipment to be used:
•
Ergometer =
•
Treadmill =
•
None (Pharmacological) = Ergom. button, then select Ergoselect, then None
•
Return to Patient Information = Back button
5. Touch each protocol field selection that you wish to modify.
•
Available choices will be displayed.
•
The keyboard will appear for numeric entries.
•
Changes are saved and available for all future cardiac stress exams.
Ergometer Protocol Settings
Select Ergometer:
Pre-Exercise Load (Watts):
Exercise Starting Load (Watts):
Exercise Stage Time (mm:ss):
Exercise Load Increment (Watts):
Exercise Auto Print:
Recovery Load (Watts):
Recovery Print Period (mm:ss):
Treadmill Protocol Settings
Select Treadmill:
Select Protocol:
Units for Speed:
Pre-Exercise Speed:
Exercise Auto Print:
Recovery Speed:
Recovery Print Period (mm:ss):
. button
button
Choices
None or Ergoselect
Any numeric value in increments of
5 from 10 to 995
Any numeric value in increments of
5 from 10 to 995
Minutes (mm): Seconds (ss) from
0:10 to 10:00
Any numeric value in increments of
5 from 10 to 995
No or Yes
Any numeric value in increments of
5 from 10 to 995
Minutes (mm): Seconds (ss) from
0:10 to 10:00
Choices
TrackMaster TMX 428
Bruce, Modified Bruce, Naughton,
or Balke
MPH or km/h
0.5 to any numeric value
No or Yes
0.5 to any numeric value
Minutes (mm): Seconds (ss)
USING THE ELI 280 EXERCISE STRESS OPTION
Default Settings
Ergoselect
10
25
3:00
25
Yes
10
3:00
Default Settings
TrackMaster TMX 428
Bruce
MPH
Protocol dependent
Yes
Protocol dependent
Protocol dependent
Page | 10