The technical specification details of Comfy Stim® are as follows:
MECHANISM
01
Channel
02
Pulse Amplitude
03
Wave Form
04
Voltage
05
Power source
06
Size
34
07
Weight
08
Pulse Rate
09
Pulse Width
10
On Time
11
Off Time
12
Ramp Time
13
Mode
TECHNICAL SPECIFICATIONS
14
Burst Mode
15
Normal Mode
16
Modulation Mode
17
SD1 Mode
Telephone: 0800 121 4626
TECHNICAL DESCRIPTION
Dual, isolated between channels
Adjustable, 0-100 mA peak into 500 ohm load each channel.
Asymmetrical Bi-Phasic Square Pulse
0 to 50V (Load: 500 ohm)
One 9 Volt Battery
10.1cm(L) x 6.1cm(W) x 2.45cm(H)
150 grams with battery.
Adjustable, from 2 to 150 Hz, 1 Hz/step
Adjustable, from 50 to 300 microseconds, 10 us/step
Adjustable, 2~90 seconds , 1 Sec./ step
Adjustable, 0~90 seconds , 1 Sec./ step
Adjustable, 1~8 seconds, 1 Sec./ step, The "On" time will increase and
decrease in the setting value.
Five TENS Modes: B(Burst), N(Normal),M (Modulation),SD1( Strength
Duration), SD2 Two EMS Modes: S(Synchronous), A (Alternate)
Burst rate: Adjustable, 0.5 – 5Hz Pulse width adjustable, 50~300µs
Frequency fixed = 100 Hz
The pulse rate and pulse width are adjustable. It generates continuous
stimulation based on the setting value.
Modulation mode is a combination of pulse rate and pulse width
modulation. The pulse rate and width are automatically varied in a cycle
pattern. The pulse width is decreased by 50% from its original setting in
0.5 second, then the pulse rate is decreased by 50% from its original
setting in 0.5 second. Total cycle time is 1 second. In this mode, pulse
ate(2-150Hz) and pulse width(50-300µs) are fully adjustable.
The SD1(Strength-Duration) mode consists of automatic modulation
intensity and pulse width in 40% range. The intensity is always increasing
while the pulse width is decreasing and vice-versa. The intensity is
decreased by 40% while the pulse width is increased by 40% in 5 seconds.
In the next 5 seconds, the intensity is increased by 40% while the pulse
width is decreased by 40%. Total cycle time is 10 seconds. Pulse rate (
2~150Hz) and pulse width (50~300µs) are fully adjustable.