k. Set Pulse Width
Pulse Width is adjustable from 50 us to 300 us. Press "SET" control to enter this
menu, then press "Increment" or "Decrement" to adjust the setting. If no
instructions regarding the pulse width are given in therapy, set the control to
the suggested 70-120 us setting.
l. Set Pulse Rate
Pulse rate is adjustable from 2Hz to 150 Hz . Press "SET" control to enter this
menu, then press "Increment" or "Decrement" to adjust the setting. Unless
14
otherwise instructed, turn the pulse rate control to the 70-120 Hz range.
m. Set Timer
The treatment time is adjustable from 1 to 60 minutes or C (Continuous).
ADJUSTING THE CONTROLS FOR EMS
Press "SET" control to enter this menu, then press "Increment" or
"Decrement" to adjust the setting. Press "Increment" control when the timer
shows 60 minutes, it will be switched to continuous stimulation.
Telephone: 0800 121 4626