Covaris LE220 Service Note

Transducer/match network replacement
Hide thumbs Also See for LE220:

Advertisement

Service Note: LE220 Transducer (and Match Network) Replacement
Purpose:
This service note provides instruction for replacement of the Transducer and Match
Network in an LE220 system.
NOTE:
The Transducer and Match Network are both serialized and must be paired together.
Upon replacement, new values must be programmed onto the RF Power Board via SonoLab.
Parts:
500322, LE220 Transducer and Match Network Repair Kit
Tools / Equipment Required
• 9/16" Open End Wrench
• 3mm Hex Wrench or Driver
• Philips Head Screwdriver (#1)
• Wire Cutters or Scissors

1. Remove Existing Transducer and Match Network

1.1.
Open Front Door to LE220. Then turn OFF LE220 Instrument and disconnect
AC Power Cord from instrument.
1.2.
Remove top cover by unscrewing 8 flathead screws on top of the instrument,
and 2 pan head screws at the upper rear. Set cover aside.
Document Number: 410053
LE220 Transducer/Match Network Replacement
Rev B
Page 1 of 6

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LE220 and is the answer not in the manual?

Questions and answers

Summary of Contents for Covaris LE220

  • Page 1 • Wire Cutters or Scissors 1. Remove Existing Transducer and Match Network 1.1. Open Front Door to LE220. Then turn OFF LE220 Instrument and disconnect AC Power Cord from instrument. 1.2. Remove top cover by unscrewing 8 flathead screws on top of the instrument, and 2 pan head screws at the upper rear.
  • Page 2 Rotate knob for manual adjustment of transducer height 1.5. Remove BNC Transducer Cable from mating connector at inner connector panel location (see image). Transducer BNC Cable Document Number: 410053 Rev B LE220 Transducer/Match Network Replacement Page 2 of 6...
  • Page 3 1.8. Use a 9/16 Open End Wrench to loosen and remove the nut (and lock washer) securing the match network BNC adapter to the inner connector panel. Document Number: 410053 Rev B LE220 Transducer/Match Network Replacement Page 3 of 6...
  • Page 4 Serial Number of the Match Network. 2.1.4. While carefully holding transducer (and brackets) in place, install 4 hex screws into manifold. Do not tighten all screws until they are all in place. Document Number: 410053 Rev B LE220 Transducer/Match Network Replacement Page 4 of 6...
  • Page 5 Secure Transducer cable in place by re-applying tie wraps (locations shown in Section 1.3). Trim excess length as needed. 2.1.7. Replace the top cover on the LE220 using the 10 screws indicated before. 3. Re-Configure LE220 3.1. Re-connect AC line cord to the instrument. Turn ON the LE220.
  • Page 6 3.10. Click the Log Out button to log out as ‘service’. 3.11. If desired, un-check the Enable Authentication box. Verify transducer settings have been changed by reviewing content in About SonoLab tab. Document Number: 410053 Rev B LE220 Transducer/Match Network Replacement Page 6 of 6...

Table of Contents