Covaris
Quick Start Guide - M220 Focused-ultrasonicator
Instrument
Unpacking and Setup:
Space requirements: Height 19 1/2" (with computer open), Width 14", & Depth 10". Two 110
electric plugs are required.
• M220 is shipped in a padded cradle inside the cardboard shipping box. Lift cradle from
the cardboard box by using the lifting handles. Remove the M220 system from the
cradle and place it on the bench top. CAUTION: Lift the M220 out of the box by its
baseplate. Do not lift by the acoustic assembly. Remove all wrapping materials.
XTU holder
with insert
Hinges
Safety
Cover
Acoustic
Assembly
• Unpack all accessories included with the instrument shipment. Items include the
Notebook computer with AC adapter, 1 USB Cable, 1 M220 Power cord, fill-drain kit,
M220 Holder XTU, inserts for various size microTUBEs, 100 ml wash bottle, 20 ml
syringe, 10 d isposable swabs, AFA grade water (125 ml), and a safety cover. Additional
items may be in the shipment if they were included in the order.
• Peel off the backing from the Velcro stripes on the computer loading bars. Place laptop
computer on the top of the M220 using the Velcro strips to secure it in place.
• Connect the laptop to the M220 using the USB cable. The port for the USB cable is on
the back on the M220.
• Plug in M220 power cord and laptop power cord.
• Place the safety cover on the front of the M220, on the hinges, and close it.
• The M220 and laptop can now be powered on. Power Switch for M220 is located on
the back as shown below.
Computer
Loading Bar
Chiller
Exhaust
(on both
sides)
PN 010161 | Rev D
www.covaris.com