To measure the harmonic ratio, set the marker on the signal and a delta
marker on the second harmonic.
9. Press [MARKER] key.
The R&S Spectrum Rider sets a marker on the trace maximum. The trace
maximum corresponds to the signal.
10. Select the "New Marker" softkey at the "Measurement footer bar".
The R&S Spectrum Rider activates a delta marker and places it on the next
trace maximum. This corresponds to the second harmonic.
The harmonic ratio is the vertical distance of the marker and the delta marker.
The R&S Spectrum Rider displays this value in the "Measurement result
view".
5.2

Using a Power Sensor

R&S Spectrum Rider option
R&S FPH-K9 (order number: 1321.0709.02) option is required to operate
the R&S Spectrum Rider in the power sensor mode.
Getting Started 1321.0996.02 ─ 09
Trying Out the Instrument
Using a Power Sensor