Optical head interface modules and optical heads (114 pages)
Summary of Contents for Keysight Technologies E8257DS Series
Page 1
Keysight Technologies E8257DS02-S15 User’s Guide Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com. Millimeter Source Modules...
The E8257DS02, S03, S05, S06, S08, S10, S12, and S15 are Millimeter-wave Source Modules that are manufactured by Oleson Microwave Labs (OML) and is provided as an ordering convenience by Keysight Technologies. This product can be ordered directly from OML. For additional Millimeter-wave Source Modules information go to the Keysight website at http://www.keysight.com/find...
E8257DS02-S15 System Setup System Setup 1. Connect the interface cable (included with this option) from the millimeter-wave source module and to the rear panel on the PSG. Refer to Figure 2. Connect the RF cable (included with this option) from the SMA RF Input connector on the millimeter-wave source module to the RF Output connector on the PSG, using the adapter listed in Table...
E8257DS02-S15 Performance Specifications Performance Specifications The following table provides specifications or each of the millimeter source models. Refer to the millimeter source module used for your application. All specification apply over a temperature range of +20 to +30 °C. Table 2 Millimeter Source Model Performance Specifications Item Multipliers...
Page 9
E8257DS02-S15 Performance Specifications Table 3 Instrument Dimensions Net Weight Wid th Height Depth < 2.0 lb (907 g) nominal 5.7 in (14.47 cm) 2.8 in (7.11 cm) 4.3 in (10.92 cm) < 4.0 lb (1.8 kg) 9.5 in (24.13 cm) Height excludes the adjustable rubber feet length and attenuator adjustment knob.
E8257DS02-S15 Performance Specifications Attenuator Options Attenuator options will slightly degrade the maximum power output from the module. A25- Manually Ad justable Attenuator • micrometer controlled • 25 dB range minimum (not available on the E8257DS02) EA2- Electronically Controlled Attenuator Available on the E8257DS12 (only). •...
Declaration of Conformity A copy of the Declaration of Conformity is available upon request, or a copy is available on the Keysight Technologies web site at https://regulations.about.keysight.com/DoC/ Statement of Compliance This instrument has been designed and tested in accordance with CAN/CSA 22.2 ...
E8257DS02-S15 Safety and Regulatory Information Before Applying Power Verify that the premises electrical supply is within the range of the instrument. The instrument has an autoranging power supply. If this product is not used as specified, the protection provided by the equipment could be impaired.
E8257DS02-S15 Safety and Regulatory Information Servicing These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
E8257DS02-S15 Safety and Regulatory Information Electrostatic Discharge Protection Protection against electrostatic discharge (ESD is essential while removing assemblies from or connecting cables to the instrument. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage.
E8257DS02-S15 Safety and Regulatory Information Instrument Markings Listed below are definitions for the markings that may be found on the product. The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation. The AC symbol indicates the required nature of the line module input power.
E8257DS02-S15 Safety and Regulatory Information Electrical Safety Compliance Complies with the standards (dates and editions are cited in the Declaration of Conformity): • IEC/EN 61010-1 Acoustic Statement: (European Machinery Directive): • Acoustical noise emission LpA<70 dB Operator position Normal operation mode Per ISO 7779 To find a current Declaration of Conformity for a specific Keysight product, go to: http://www.keysight.com/go/conformity EMC Compliance...
Shipping Your Prod uct to Keysight for Service or Repair If you wish to send your instrument to Keysight Technologies for service or repair: • Include a complete description of the service requested or of the failure and a description of any failed test and any error message.