DataLogs and Event Logs

DataLogs
The detector records various information to create a report. The detector is capable of storing
45 days of data under the following condition estimation:
8 hours and 2 minutes of alarm per day.
l
A recording of data logs in 15 seconds interval.
l
Power ON/OFF twice per day.
l
One Bump Test per day
l
When the memory is full, the detector replaces the oldest datalogs with the most recent
datalogs.
Event Logs
The detector records a maximum of 50 gas alarm and maintenance events and error conditions.
The following alarm events are recorded:
Low alarm
l
High alarm
l
STEL alarm
l
TWA alarm
l
Over range
l
Negative
l
The following maintenance events are recorded:
Automated bump test or calibration
l
Manual sensor zero
l
Bump test
l
Calibration
l
Any error condition is recorded, such as:
BLE not found
l
RTC fail
l
Broken Datalog link
l
Sensor fail (reflex, count out of range, no communication)
l
Temperature Fail
l