10. Open the cylinder valve by turning the pressure regulator knob clockcounterwise. The Zero
process starts and a message is displayed when succeeded.
11. Follow onscreen instructions to know when to apply gas and when the calibration process is
complete.
12. The process is complete when the results are displayed on your mobile device. You can now
remove the cap by pulling on the tabs.

Bump Test

The detector can be tested in two ways:
• Apply gas from a cylinder to the sensors manually through the calibration cap, and using the
Safety Suite Device Configurator (SSDC) software or the Device Configurator (DC) app.
• Use an IntelliDoX module.
Move to a normal atmosphere (20.9% v/v O
test gas.
Details for Bump Test and maintenance:
Recommendations for initial checking of the equipment on a routine basis including the
l
maximum time interval between calibrations.
Perform a functional check with gas before each day of use.
l
Honeywell recommends bump testing the sensors before each day's use to confirm their
l
ability to respond to gas by exposing the apparatus to a gas concentration that exceeds the
alarm setpoints. Manually verify that the audible and visual alarms activate.
The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different
l
combustible gas in the % LEL range, calibrate the sensor using the appropriate gas.