Nulling Electrical Offsets
Temperature and humidity variations affect the performance of electronic
circuits and optical detectors. Nulling the electrical offsets eliminates these
effects.
If light reaches the detector when nulling offsets, you will be
notified and the nulling is not performed. You can retry after
putting a cap on the detector.
To perform an offset nulling:
1. If you have not done so already, put a cap on the detector.
2. From the main PX1 page, tap the
select
3. Under Power Meter, tap Offset nulling.
4. Tap Nulling and wait for the operation to complete. Once it is done, tap
OK to return to the Settings page.
Optical Power Expert
I
MPORTANT
.
Setting up and Using Your PX1
Nulling Electrical Offsets
icon to access the menu, then
31