Performance Verification And Adjustments; Equipment List - Keysight Technologies U2020 X Series Service Manual

Usb peak and average power sensors
Hide thumbs Also See for U2020 X Series:
Table of Contents

Advertisement

2

Performance Verification and Adjustments

Equipment List

The following equipment are required for performance verification:
Equipment
Network analyzer
Calibration kit
Signal generator
Power meter
Power sensor
Wideband power sensor
Cable adapter
Power splitter
Pulse pattern generator
Frequency counter
20
Critical specification
Frequency range: 10 MHz to 40 GHz or above
Dynamic range: 94 dB
Trace noise: <0.006 dB
Frequency range: DC to 18 GHz or above
Power range: –50 dBm to +22 dBm at 1 GHz
Output resistance: 50 W
Dual-channel peak power meter, compatible with
P-Series and N8480 Series power sensors
Absolute accuracy: ±0.8%
Frequency: 50 MHz or above
Power range: –30 dBm to +20 dBm
Standing wave ratio (SWR): £1.15 at 50 MHz
Power range: –30 dBm to +20 dBm
SWR: £1.20 at 1 GHz
11730A sensor cable adapter
Two-resistor type power splitter, N-type (f)
Maximum frequency: 18 GHz
Pulse pattern generator that can generate <12 ns
pulse period
Frequency counter that can count up to 20 MHz
frequency
Recommended Keysight model
number/part number
E8361A/E8361C/E8363B/E8363C
85054A/85054D/85056A/85056D
E8257D/N5182A/E4438C
N8481A/N8487A
N1921A/N1922A
N1911A-200
Keysight U2020 X-Series Service Guide
N1912A
11667A
81130A
53132A