Cleaning
Use a clean, water-dampened cloth to clean the body of the U2020 X-Series.
Connector care
A solution of pure isopropyl or ethyl alchohol can be used to clean the connector
but make sure to keep in mind on its flammable nature.
– The RF connector beads deteriorate when contacted by hydrocarbon
CAUTION
– Do not attempt to clean the connector with anything metallic such as
– Clean the connector only at a static-free workstation. Electrostatic
Clean the connector face by first using a blast of compressed air. If the
compressed air fails to remove contaminants, use a foam swab dipped in isopropyl
or ethyl alcohol. If the swab is too big, use a round wooden toothpick wrapped in a
lint-free cloth dipped in isopropyl or ethyl alcohol.
Keysight U2020 X-Series Service Guide
compounds such as acetone, trichloroethylene, carbon tetrachloride, and
benzene.
pins or
paper clips.
discharge to the center pin of the connector will render the U2020
X-Series inoperative.
General Information