53132A SCPI commands
*RST
*CLS
INP1:IMP 50
INIT
FETC?
Procedure
1 Ensure that the DUT has been allowed to warm up for approximately an hour
2 Send the commands as provided above. The frequency from the 53132A must
3 If the verification fails, refer to
Keysight U2020 X-Series Service Guide
before attempting to measure timebase accuracy.
be within 10 MHz ± 250 Hz.
Performance Verification and Adjustments
"Adjustments"
on page 36.
2
35