Taring/Net Compensation - Vetter Smartfork Operating Instructions Manual

Scale weighing fork arms
Hide thumbs Also See for Smartfork:
Table of Contents

Advertisement

Fig. 12: Calibration points, SmartFork
1
Rear calibration point
2
Front calibration point
The following is required for calibration:
n
Lifting device for large loads, e.g. indoor crane.
n
Have a second person on hand.
n
A calibration weight weighing at least 100 kg which is attached to 2 points close to the
load cell.
n
A reference weight weighing roughly 60% of the rated load of the weighing fork arm.
The mass of the reference must be known with the greatest possible accuracy.
For calibration, raise the SmartFork
the calibration weight to the defined positions (Fig. 12).
5.1.3

Taring/net compensation

V1.02/2020
Weight determination with the SmartFork® Scale > Taring/net compensation
®
Timeout
If the SmartFork
®
Scale remains in calibration mode for longer than 5
minutes without a weight being attached, the system on the SmartDis-
play switches itself off.
Operating manual, SmartDisplay
The complete manual and all steps for taring can be found in the oper-
ating manual for the SmartDisplay in the other applicable documents.
®