ITech IT8500plus Series User Manual page 59

Dc programmable electronic loads
Hide thumbs Also See for IT8500plus Series:
Table of Contents

Advertisement

Set the power-on state to automatic test
1.
Press [Shift] + [8] (System) to enter the system menu.
POWER-ON BUZZER
2.
Press the right key to the RUNMODE, and press [Enter].
RUNMODE DEFAULT
3.
Press the left or right key to select PROG_TEST and press [Enter], the
menu will return to RUNMODE.
RUN <PROG_TEST
4.
Press [Esc] to exit.
P01
Recall the test file to run
The following method can quickly recall the existing test file from the EEPROM
and test it after the instrument is powered on.
Operate in automatic test mode:
1.
Press [Shift] + [Enter] (Recall).
0.0000V
RECALL
2.
Press the number keys to select the group and press [Enter].
0.0000V
P01
3.
If the start voltage (Start) is set to 0 during the editing process, press [Shift]
+ [.] (Trigger) to trigger.
If the start voltage (Start) is set to a non-zero value, connect the DUT to
start the test. The automatic test step will be displayed on the VFD and
FAULT or PASS will be displayed after the test is over.
0.0000V
P01-01
Note
To exit the automatic test mode, press [Shift] + [8] (System) to enter
the system menu and select RUNMODE as NORMAL.
0.000A
PROGRAM=1 (0~10)
0.000A
0.000A
2.100A
Note
Press [Shift] + [0] (Pause) to stop running the automatic test. Press
to continue testing.