Table 4 – Guidance and manufacturer's declaration – electromagnetic immunity – for
all medical electrical equipment and medical electrical systems that are not life-
supporting
Guidance and manufacturer's declaration – electromagnetic immunity
The B.O.L.T device is intended for use in the electromagnetic environment specified below. The customer or
the user of the B.O.L.T device should assure that it is used in such an environment.
Immunity test
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
© Copyright 2019, All Rights Reserved. AmZetta Technologies Private Ltd.
60601 test level
3 Vrms
3 Vrms
150 kHz to 80 MHz
150 kHz to 80
6 V in ISM and
MHz
amateur
radio
6 V in ISM and
bands
between
amateur
0.15 MHz and 80
bands between
MHz
0.15 MHz and
80% AM at 1 kHz
80 MHz
80% AM at 1
kHz
10 V/m
80 MHz to 2,7 GHz
80 % AM at 1kHz
10 V/m
80 MHz to 2,7
GHz
80 % AM at
1kHz
3 V/m
2.7 GHz to 6
GHz
Proximity
fields
80 % AM at
from RF wireless
1kHz
communications
equipment (As per
table 9)
Proximity fields
from RF wireless
communications
equipment (As
per table 9)
Electromagnetic environment –
Compliance
level
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of B.O.L.T device, including
cables,
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1,2√P 150kHz to 80 MHz
d = 0.6√P 80 MHz to 2,7 GHz
radio
d = 2,3√P 80 MHz to 6 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according
manufacturer and d is the recommended
separation distance in meters (m).
Field
transmitters, as determined by an
electromagnetic site survey should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following