Page 1
Procedures requiring work inside the indicator must be performed by qualified service personnel only. The supply cord serves as the power disconnect for the 880 Analog Output Installation. The power receptacle to the indicator must be easily accessible.
Page 2
Also ensure SW1 located on the front of the card is in the OFF position and SW3 and SW4 located on the back of the board are in the OFF positions. 880 Analog Output Installation...
Page 3
on off Back Front SW3/ 4 Figure 3. Switch Locations on Analog Output Card 8. Install the Analog Output card on the CPU board standoffs, plugging it into the J8 connector on the CPU board (Figure 4). 9. Use screws to secure the option card and face plate to the supporting standoffs (Figure 4). Analog Option Faceplate Analog Output Card...
Page 4
Before securing the backplate, verify the display connector aligns properly with the front cutout. If not, remove the boards and backplate and re-insert with the enclosure in another position so that everything does line up. Figure 6. Boards Installed in Indicator Assembly Enclosure 880 Analog Output Installation...
Page 5
The indicator automatically recognizes all installed option cards when the unit is powered on. No hardware- specific configuration is required to identify the newly-installed card to the system. 880 Universal Instructions 1. Disconnect power to the indicator. 2. Open the indicator enclosure to access the CPU board.
Page 6
If the zero value read on the multimeter is correct, press to cancel the re-calibration. Otherwise, re- MENU calibrate as in Step 3 above. • Press to select the digit • Press to increment or decrement the value • Press TARE when the value is correct 880 Analog Output Installation...
Page 7
If recalibrating the zero point, also re-check the span point and calibrate if needed. Repeat until both readings are correct. 6. Return to weigh mode. Analog output function can be verified using test weights. See the 880 Technical Manual (PN 158387) for more configuration information. Specifications Resolution:...
Page 8