Hach Polymetron 9611sc PO43-LR Operation Manual page 12

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
Option
SAMPLE MISSING
SET DATE & TIME
DISPLAY SETUP
DISABLE REMINDERS
MANAGE DEVICES
INSTRUMENT
INFORMATION
CALCULATION
SETUP OUTPUTS
SETUP NETWORK
(conditional)
SECURITY SETUP
AIR PURGE
RESET DEFAULTS
Manage devices
Install or remove input modules.
1. Push menu and select MANAGE DEVICES.
2. Select an option.
Option
SCAN FOR DEVICES The system shows the connected devices.
DELETE DEVICE
Set up the calculation
Set up variables, parameters, units and formulas for the analyzer.
1. Push menu and select CALCULATION.
2. Select an option.
Option
SET VARIABLE X
12 English
Description
Sets the operation that occurs if no sample is detected. Options: ON DELAY
(default) or OFF DELAY. ON DELAY: the instrument waits the remainder of the
measurement cycle, then goes to the next sample in the sequence. OFF DELAY:
the instrument waits 10 seconds, then goes to the next sample in the sequence.
Sets the analyzer time and date.
Changes the language. Adjusts the order measurements are shown. Adjusts the
display contrast settings.
Stops scheduled maintenance alerts for individual components. Options: tubing,
stir bar, colorimeter cell, air filter, stir motor, air relief valve, pinch valve, air
compressor, reagent valves, sample valves, standard valves, colorimeter LED, fan
filter, air check valve.
Installs or removes input modules. Refer to
more information.
Shows the analyzer information. Refer to
on page 18.
Set up variables, parameters, units and formulas for the analyzer. Refer to
the calculation
on page 12.
Selects and configures the 4–20 mA setup, the relay setup and the error hold
mode. Refer to
Configure the outputs
Shows only if a network card is installed. The network card support is for Modbus,
Profibus and HART.
Enables or disables the pass code (default = HACH55).
Enables the air purge for use with an external air supply. Options: on or off
(default). Off: an external air supply is not used. The vent fan is enabled and the
air filter is installed. On: an external air supply is connected to the instrument. The
vent fan is disabled. The air filter is replaced with a fan filter plug. To use this
feature, make sure that the fan filter plug is installed. Refer to the documentation
that is supplied with the air purge kit.
Sets the configuration to the factory defaults.
Description
Note: If no device is connected, the system returns to the main measurement
screen.
Removes the device when the device is no longer connected.
Description
Selects the sensor referring to the variable X.
Manage devices
on page 12 for
View instrument information
on page 13 for more information.